This walkthrough will help you to re-start the pumps in the molecular beam instrument at US5.
When power is out
- In case of a power outage: Check if the power is still down, and check the pressures on the MaxiGauge on the instrument (control panel on the right side of the instrument).
When power comes back
- Check the two fore vacuum pumps under the stairs: the scroll pump should start automatically when power comes up, the rotary vane will stay off (keep it that way).
- Ensure that the fore vacuum line is only open toward the intracavity turbo pumps.
- Switch on the main power of the Bestec 19" rack.
- Ensure pumps P1, P2 and P3 start have started automatically (they should show yellow, indicating a rising pump speed).
- If not: start them by left-clicking either of the pumps and select "start" in the popped-up window; Then close the window.
- (Usually, starting one will also start the other two, but this sometimes fails. To be sure, start each one explicitly).
- (If pump stays red, displaying 0 Hz, communication may have failed. Cycling the power supply frequenly helps).
- If you have not switched off the power supplies (in the "when power is out section"): stop pumps P4, P5, P6 and P7.
- Pressures on G1, G2 and G3 should rapidly drop to 10-5, the foreline pressure is G8, it should be 10-2—0.1 mbar
- Contact Joost at +31 6 4412 6316 or Olga at ...