This walkthrough will help you to re-start the pumps in the molecular beam instrument at US4.
When power is out
- In case of a power outage contact Sjors at +31 6 23848434 or Sander at +31618875389
The setup should be filled with nitrogen
- If power is back up, the backing pumps should be running and the setup should be pre-pumped
- Check the MEMB pressure
- if MEMB pressure is several mbar, you can turn on the turbo pumps
- first turn on the big turbopump using the on/off button on the top right controller (DCU 600)
- then turn on the other turbo pumps using the two bottom controllers:
The pfeiffer pump (bottom left) can be turned on by simply pushing the on/off button The edwards pump (bottom right) can be turned on by