Dynamic Cavity Desynchronization (DCD)
DCD with the AWG no additional boxes required

  1. Select ACPS 80 output 8 = "general purpose"
  2. Connect ACPS 80 met M22
  3. Use Timer AWG unit 2, AWG Chanel 1 to adjust the delay and amplitude
-what worked was:
delay between 20 and 50
Δλ scales linearly with amplitude Amplitude between 60 and 80%
  1. set the cavity length between 0.7 and 1.1 λ-detuning
  2. After the machine is properly set up, play with the delay, amplitude and cavity length for best results. This is the flatest possible spectrum and the highest possible power.

(9 mJ @16 µm met een 2.1mm gat)