From HFML-FELIX Wiki
Jump to: navigation, search

Tuning the Electron Source Resonator

[edit | edit source]

Grid Pulser Taper

[edit | edit source]
Cathode Grid Pulser

The Grid pulser should be fitted to the FLARE Cathode if you wanna drive it in 20MHz mode.




















3GHz Resonator

[edit | edit source]
3Ghz Resonator fitted on Cathode

When the Resonator is put back onto the Cathode don't forget to connect the RF input cable, and turn back on the RF amplifier in the GTA afterwards.




















Preparing the Network Analyzer

[edit | edit source]
Calibration kit


Ready for calibration


  1. Heat the electron source to red heated.
  2. Turn on the network analyzer (now called NA).
  3. Connect the calibration kit via the usb connector.
  4. Wait for the calibration kit to be ready, red LED turns green after approximately 15mins.
  5. Connect Port 1 cable to PORT A on the calibration kit.
  6. Connect Port 2 cable to PORT B on the calibration kit.
  7. Set Stimilus Center at 2,9979245800 [Ghz]
  8. Set Stimilus Span at 100 [Mhz]
  9. Set Sweep setup to 1601 [points]
  10. Click on avg and set IF bandwidth to 100 [Hz]
  11. Push Calibration on the NA.
  12. Choose Ecal
  13. Do a 2 port calibration.













Measuring the FLARE Resonator.

[edit | edit source]
Measurement 100 MHz span(March 2020)
5 MHz span(March 2020)
  1. Port 1 connects to the input of the electron source.
  2. Port 2 connects to the probe of the electron source.
  3. Choose S11 measurement
  4. Set Scale to autoscale
  5. Push marker search and choose min
  6. Turn on auto update, so each time a measurement is done it tracks the minimum.
  7. Now turn on Marker 2 and set this at the position of 2,9979245800 [Ghz]
  8. Now turn the pulser board ring until marker 1 is at the same position as marker 2. ( it is wise to increase the IF bandwidth to 2[Khz] when the frequency is way off)