At the end of 2020, the attenuator box at the flare diagnostic station was made to be remotely controlled. Mechanics project P002,(at the moment also a folder at ffg2), has the following specifications:

  • The box has 6 different attenuator grids that dim the incoming laser light: 3 dB, 5 dB, 10 dB, 10 dB and 10 dB.
  • Operation air pressure: 6 bar.
  • Estimating moving mass max: 0,5 kg per grid.
  • Cilindertype: DSNU-8-10-P-A (FESTO), 8 mm cilinder, extimated force 18 N, stroke 100 mm. See file: Data Sheet DSNU cilinder
  • Position sensor: SMT-8M-A-NS-24V-E-2,5-OE (FESTO), magnetic reed sensor.
  • Air valve manifold: VTUG-10-SH2-B1T-Q6L-U-Q6S-5J+H. See file: Offerte Ventiel Eiland en meer, Festo 25 mei 2020 [
  • The frontplate was replace to fit cilinder. See file: Front Plate Design