| Step | Where can I find that? | Action | Details |
|---|---|---|---|
| 1 | Main panel | Switch on all 5 FELIX & Felice buttons |
|
| 2 | FELIX User Station (wnd ) | Select the correct user station |
|
| Select 25 MHz or 1 GHz | |||
| 3 | FELIX_dump_line_L2 (wnd) | Find or Load a previous dump-line setting for E-range |
|
| 4 | Find or Load the same setting in low-lever RF phases | ||
| 5 | Magneetslingercircuit (wnd) | Change current to match 2B20-25 | |
| 6 | Gun > Felix (wnd) | Change gun bias to ~170.5 V @ 25 MHz or ~80 V @ 1 GHz. | Precise number can be found in the file name |
| 7 | HPRF.vi (wnd) | Check the file has correctly loaded in the HPRF window. Set values to match the yellow values | |
| 8 | Main panel | Switch the “3GHz RF timing on†button on | |
| 9 | low level RF phases (wnd) | Set klystron modulators to target values (from file name). Step is ï¾4kV - 5kV. Press ‘feedback’ off, and ‘zero’. | |
| 10 | on Flare interface computer | If FLARE solenoids are off, switch them on. |
|
| 11 | LLRF power (wnd) | press ‘feedback’ off, and ‘zero’ all. | |
| 12 | Felix Timers (wnd) | Check timer S1 = 2µs and rep. rate = 1Hz. Open vacuum system. | |
| 13 | OC Control (wnd) | Insert BS, and align cavity mirrors. Watch 2S21, align downstream. Watch 2S41, align upstream. Remove BS. |
|
| 14 | Main panel | Press ‘GUN RF timing on’. | |
| 15 | Felix dump line L2 (wnd) | Optimize e- path. Align with indicator. PM10 = SX10. PM20 = SX11. PM30 = SX30* (fault with blue, align 1 square below indicator). PM40 = SX40. PM50 = SX55. PM60 = SX70. Slowly ramp to ï¾10µs / 10Hz, ensure all e- are dumped. | |
| 16 | low level RF phases (wnd ) | Fine-tune low level RF phases. Once done, press ‘zero’, ‘load’, and ‘feedback on’ in all. | |
| 17 | At diagnostic station | switch amplifier to x1 (@1GHz) or x10 (@25MHz). | |
| 18 | Gun / prebuncher Felix (wnd) | Adjust gun bias to achieve 360mV @25MHz / 200mV @1GHz (yellow curve on scope) | |
| 19 | OC control Felix (wnd ) | Find cavity zero. Set ‘dt_CLdetuning’ to zero, and adjust “DT_CLzero†close to an unstable state. Then, reset ‘dt_CLdetuning’ to 1.8. | |
| 20 | uc-flx-gui (wnd) | Calibrate λ. Go to calc, and set lambda to correct live value. |