MIOO = Measure-It Object Oriented
[edit | edit source]This program is a homebuilt data acquisition program, built in Labview2017 that can be used to acquire scope traces while scanning external devices, such as the FELIX/FELICE wavelength, or certain delay generators.

Source code
[edit | edit source]Source code is available at https://gitlab.science.ru.nl/jmbakker/mioo
Note that it is compiled using Labview 2017, 32 bits. Do NOT use newer versions, as that will bream compatibility with older hardware.
Available instruments for scanning
[edit | edit source]Bestec
[edit | edit source]Variation of the vertical or horizontal position of the entire Felice Molecular beam instrument. Communication via network, governed by entry in the *Instruments\Bestec.ini* file:
[Bestec] address=192.168.114.13 port=20001
| maximum value | description | |
|---|---|---|
| Horizontal | +100 | Felice Molbeam out-of-focus |
| -183 | Felice Molbeam in-focus | |
| Vertical | -32.9 | |
| -22 | ||
| safe values | ||
| Vertical | -30 | molecular beam overlaps with FELICE |
NSDL
[edit | edit source]NarrowScan dye laser, manufactured by Radiant Dyes
DG535
[edit | edit source]Stanford research systems delayg generator, model 535. Communication via GPIB.
[DG535] address=1:15
FELIX
[edit | edit source][FELIX] # save_address=isolator-private2.net.science.ru.nl OLD #address=auto address=isolator save_port_ds=18020 port_ds=18000 port_uc=2706 #werkte bij victor #port_uc=18040 port_oc=2708
[FELICE] # srfel.science.ru.nl OLD! address=isolator port_uc=3003 port_oc=2709
LaserVision OPO
[edit | edit source][OPO] address=192.168.114.90 port=10001
Lioptec
[edit | edit source]Lioptec dye laser
[Lioptec] address=192.168.114.5 port=65510
None
[edit | edit source]Essentially a do-nothing scan: take data while varying nothing, or varying something by hand
QC9520
[edit | edit source]The quantum Composers delay generator. Should be connected to a serial port, and an entry in the *Instruments\QC9520.ini* file:
[QC9520] COM=COM19
Known issues
- all values in seconds
Stepper =
[edit | edit source]A stepper motor attached to the FELICE Molbeam instruments that drives a slit in and out of the flight tube, thereby screening the ions that have not interacted with FELICE. Hardware built at Rijnhuizen projects 1795 / 1842. Currently hooked up to the fourth axis. Communication via network, governed by entry in the *Instruments\Stepper.ini* file:
[Stepper] address=192.168.114.99 port=10001 local_port=23 default_axis=4
| maximum value | description | |
|---|---|---|
| In | +14000 | completely in |
| Out | 0 | set this manually |
| safe values | ||
| Overlap | 13000 | molecular beam overlaps with FELICE |
MIOO-Analysis
[edit | edit source]Postprocessing software
Installation instructions
[edit | edit source]Prerequisites:
- a working Labview 2017 (32 bit) installation or the Labview 2017 (32 bit) runtime engine
- Downloadable at ni.com
\\hfmlfelxidisk\share\software\LabVIEW\flare-ni\Archive\LV2017\LabVIEW Runtime Installer x86
- support for hdf5 v 1.8.18
- installer at \\hfmlfelixdisk\data\cmp\clusters\labview-programs\hdf5-1.8.18
MIOO and MIOO-Analysis executables:
\\hfmlfelixdisk\data\cmp\clusters\labview-programs\mioo\mioo-v240404
\\hfmlfelixdisk\data\cmp\clusters\labview-programs\mioo-analysis\MiooAnalysis191010